Communications - Thermo Scientific iQ Series Instruction Manual

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Operation
Settings

Communications

3-90
42iQTL Instruction Manual
The Communications screen allows the user to set TCP/DHCP
parameters, Serial settings, Analog I/O and Digital I/O, Email Server, and
Instrument ID. Buttons are grayed out if not selected in
Settings>Configuration.
Home Screen>Settings>Communications
The Communications screen contains the following information:
Wired TCP/DHCP: Settings for communicating with the instrument
through wired Ethernet.
Serial RS-232/485: Settings for communicating with the instrument
through RS-232/485 protocol. This is only visible if selected in
Settings>Configuration>Communications Board.
Analog I/O: Settings for communicating with the instrument through
analog I/O settings. This is only visible if selected in
Settings>Configuration>Analog I/O.
Digital I/O: Settings for communicating with the instrument through
digital I/O settings. This is only visible if selected in
Settings>Configuration>Digital I/O.
Email Server (SMTP): Settings for communication with email.
Instrument ID: Allows the user to edit the instrument identification
number (ID). The ID is used to identify the instrument when using
protocols to control the instrument or collect data. It may be necessary
to edit the ID number if two or more of instruments of the same model
are connected to one computer. Valid instrument ID numbers are from
0 to 127. The 42iQTL has a default instrument ID of 1.