Thermo Scientific iQ Series Instruction Manual page 52

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Operation
Calibration
Adjust Span Coefficients
3-26
42iQTL Instruction Manual
The Adjust Span Coefficients screens are used to manually adjust the span
coefficients.
The following screen is shown in single range mode and dual or auto range
mode. In dual or auto range modes, "High Range" or "Low Range"
buttons are displayed to indicate the calibration of the high or low
coefficient. The Adjust High Span Coefficient and Adjust Low Span
Coefficient screens function the same way as the (single range) Adjust Span
Coefficient screen.
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (single range mode)
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Span Coefficients (dual range mode)
The Adjust Span Coefficients screen contains the following information:
Calibrate NO Coefficient: Adjusts the NO coefficient when in single
range mode.
Calibrate NO High Range Coefficient: Adjusts the NO high range
coefficient when in dual or auto range mode.