Operation
Settings

Communications

3-84
49iQ Instruction Manual
The Communications screen allows the user to set TCP/DHCP
parameters, Serial settings, Analog I/O and Digital I/O, Email Server, and
Instrument ID. Buttons are grayed out if not selected in
Settings>Configuration.
Home Screen>Settings>Communications
The Communications screen contains the following information:
Wired TCP/DHCP: Settings for communicating with the instrument
through wired Ethernet.
Serial RS-232/485: Settings for communicating with the instrument
through RS-232/485 protocol. This is only visible if selected in
Settings>Configuration>Communications Board.
Analog I/O: Settings for communicating with the instrument through
analog I/O settings. This is only visible if selected in
Settings>Configuration>Analog I/O.
Digital I/O: Settings for communicating with the instrument through
digital I/O settings. This is only visible if selected in
Settings>Configuration>Digital I/O.
Email Server (SMTP): Settings for communication with email.
Instrument ID: Allows the user to edit the instrument identification
number (ID). The ID is used to identify the instrument when using
protocols to control the instrument or collect data. It may be necessary
to edit the ID number if two or more of instruments of the same model
are connected to one computer. Valid instrument ID numbers are from
0 to 127. The 49iQ has a default instrument ID of 1.
Bayern Hessen Settings: User selects the registers (measured values) that
the instrument should respond with, in response to a valid data query
(DA) command.