Thermo Scientific iQ Series Instruction Manual page 203

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Thermo Scientific
Problem
Alarm – NO Bkg Check
Offset
Alarm – NO Span Check
Offset
Alarm – Ambient
Thermistor
Alarm – Chamber Pressure
Alarm – Bench
Temperature Short
Alarm – Bench
Temperature Open
Alarm – Cooler Current too
Low
Alarm – Cooler Current too
High
Alarm – Cooler Voltage too
Low
Alarm – Cooler Voltage too
High
Possible Cause
Action
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument background
Recalibrate the instrument.
calibration failed
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument span
Recalibrate the instrument.
calibration failed
Defective
Replace defective DMC board as
Flow/Pressure board
needed.
Defective
Replace defective DMC board as
Flow/Pressure board
needed.
Cable connection
Check the cable connection from
the heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the
heater.
Defective heater
Replace heater as needed.
Defective DMC board
Replace defective DMC board as
needed.
Cable connection
Check the cable connection from
the heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the
heater.
Defective heater
Replace heater as needed.
Defective DMC board
Replace defective DMC board as
needed.
Defective cooler
Replace cooler as needed.
Defective DMC board
Replace DMC board as needed.
Defective cooler
Replace cooler as needed.
Defective DMC board
Replace DMC board as needed.
Defective cooler
Replace cooler as needed.
Defective DMC board
Replace DMC board as needed.
Defective cooler
Replace cooler as needed.
Troubleshooting
Troubleshooting Guide
42iQTL Instruction Manual
6-5