PMT Cooler Board
Replacement
Thermo Scientific
Figure 7–31. Remove PMT Cooler
Use the following procedure to replace the PMT cooler board.
Equipment Required:
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove DMC Converter as described in
Removal" in this chapter.
3. Unplug all electrical connections from the board.
4. Using a #2 Phillips drive, unfasten six #6-32 pan head screws.
5. Pull board outwards.
DMC PMT Cooler and Reaction Chamber Replacement
"PMT Cooler Shroud
42iQTL Instruction Manual
Servicing
7-31