Thermo Scientific iQ Series Instruction Manual page 51

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Thermo Scientific
Home Screen>Calibration>Advanced Calibration>Manual
Calibration>Adjust Backgrounds>Adjust Prereactor Background
The Adjust NO, NO
, and Prereactor Background screens contains the
x
following information:
Adjust NO Background: User manually adjusts zero NO background.
Adjusted NO Concentration: Read only. Displays what the NO
concentration value will become when the calibrate button is pressed.
Current NO Concentration: Read only. Displays the current NO
concentration.
Adjust NO
Background: User manually adjusts zero NO
x
Adjusted NO
Concentration: Read only. Displays what the NO
x
concentration value will become when the calibrate button is pressed.
Current NO
Concentration: Read only. Displays the current NO
x
concentration.
Adjust Prereactor Background: User manually adjusts zero Prereactor
background.
Adjusted Prereactor Concentration: Read only. Displays what the
Prereactor concentration value will become when the calibrate button is
pressed.
Current Prereactor Concentration: Read only. Displays the current
Prereactor concentration.
Calibrate: When pressed, updates the background.
Operation
Calibration
background.
x
x
x
42iQTL Instruction Manual
3-25