Thermo Scientific iQ Series Instruction Manual page 55

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Thermo Scientific
Home Screen> Calibration>Advanced Calibration>Manual
Calibration>Adjust Coefficients>Adjust High Range NO Span
Coefficient (dual or auto range mode)
The Adjust NO Span Coefficient screens contain the following
information:
Adjust NO Span Coefficient: User manually adjusts NO span coefficient
when in single range mode.
Adjusted NO Concentration: Read only. Shows adjusted NO
concentration based on adjusted NO span coefficient when in single
range mode.
Current NO Concentration: Read only. Shows current NO
concentration when in single range mode.
Adjust High Range NO Span Coefficient: User manually adjusts the high
range NO span coefficient when in dual or auto range mode.
Adjusted High Range NO Concentration: Read only. Shows adjusted
high range NO concentration based on adjusted high range NO span
coefficient when in dual or auto range mode.
Current High Range NO Concentration: Read only. Shows current high
range NO concentration when in dual or auto range mode.
Adjust Low Range NO Span Coefficient: User manually adjusts the NO
low span coefficient when in dual or auto range mode.
Adjusted Low Range NO Concentration: Read only. Shows adjusted NO
concentration based on adjusted low range NO span coefficient when
in dual or auto range mode.
Adjusted Low Range NO Concentration: Read only. Shows current low
range NO concentration when in dual or auto range mode.
Calibrate: Calibrates span coefficient by saving the newly adjusted span
coefficient.
Operation
Calibration
42iQTL Instruction Manual
3-29