Servicing
DMC Converter Replacement
Converter Assembly
Board Replacement
7-40
42iQTL Instruction Manual
Use the following procedure to replace the converter assembly board.
Equipment Required:
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove DMC Converter as described in
Replacement" in this chapter.
3. Unplug electrical J6 and J7.
4. Using a #2 Phillips drive, unfasten four 6-32 pan head screws.
Figure 7–41. Converter Assembly Remove bd
5. Pull converter assembly board outwards.
6. Replace converter assembly board and assemble in reverse order.
"DMC Converter