Reaction Chamber
Cleaning and/or
Thermo Scientific
Use the following procedure to clean or remove the reaction chamber.
Equipment Required:
Removal
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Remove cooler/reaction chamber DMC as described in
Shroud
3. Unplug heater J8 from board.
Figure 7–33. Photo Reaction Chamber Cleaning, Electrical
4. Disconnect plumbing as shown.
5. Using a 9/64-inch hex drive, unfasten three #8-32 socket cap head
screws.
Removal" in this chapter.
DMC PMT Cooler and Reaction Chamber Replacement
"PMT Cooler
42iQTL Instruction Manual
Servicing
7-33