DMC PMT Cooler
and Reaction
Chamber
Replacement
PMT Cooler Shroud
Thermo Scientific
Use the following procedure to replace the PMT cooler and clean or
replace the reaction chamber assembly.
Remove the PMT Cooler
Replacing the PMT board
Removing and Cleaning the Reaction chamber
Replacing the PMT Cooler and Reaction chamber
PMT Replacement
Use the following procedure to remove the PMT cooler from the
instrument.
Removal
Equipment Required:
Phillips drive, #2
1. Turn the instrument OFF, unplug the power cord, and remove the
cover
Icon Here
Make sure the PMT has cooled down before touching.
2. Unhook plumbing by pulling loop downwards and unhook tubing.
3. Using a #2 Phillips drive, unfasten two 6-32 flat head screws.
4. Pull shroud outwards.
(Figure
2–1).

DMC PMT Cooler and Reaction Chamber Replacement

42iQTL Instruction Manual
Servicing
7-29