Thermo Scientific iQ Series Instruction Manual page 74

Trace level no-no2-nox analyzer
Hide thumbs Also See for iQ Series:

Advertisement

Operation
Settings
3-48
42iQTL Instruction Manual
Chamber Temp Sensor Open: Displays OK/Fail for chamber
temperature sensor open.
Cooler Temperature: Displays OK/Fail for cooler temperature. User
can adjust low and high alarm limits.
Cooler Temp Sensor Short: Displays OK/Fail for cooler temperature
sensor short.
Cooler Temp Sensor Open: Displays OK/Fail for cooler temperature
sensor open.
Cooler Current: Displays the cooler current reading.
Cooler Current too Low: Displays OK/Fail for cooler current too
low.
Cooler Current too High: Displays OK/Fail for cooler current too
high.
Cooler Voltage: Displays the current cooler voltage reading.
Cooler Voltage too Low: Displays OK/Fail for cooler voltage too low.
Cooler Voltage too High: Displays OK/Fail for cooler voltage too
high.
PMT Voltage: Displays the current PMT voltage reading.
PMT Voltage too Low: Displays OK/Fail for PMT voltage too low.
PMT Voltage too High: Displays OK/Fail for PMT voltage too
high.
Frequency: Displays the current frequency reading.
Frequency too Low: Displays OK/Fail for frequency too low.
Frequency too High: Displays OK/Fail for frequency too high.
Low Gain: Displays the current low gain reading.
Gain: Displays the current gain reading.
Board Communication: Displays OK/Fail for board communication
status.
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
3.3 V Diagnostic: Displays current voltage readings. Alarm
limits are not changeable.
5 V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.