Analog Output Zero
Thermo Scientific
Use the following procedure to calibrate the output channel to low level.
This analog output calibration procedure reflects the zero calibration for
Calibration
analog output voltage for demonstration purposes. To calibrate the 4 mA
current calibration, follow the same procedure, by selecting the 4 mA
current calibration option.
Note This adjustment should only be performed by an instrument service
technician.
1. From the Home screen, choose Settings>Communications>Analog
I/O>Analog Out Calibration.
2. Depending on the output type being used, select either Analog Out
Zero Calibration (Voltage) or Analog Out 4.000 mA Calibration
(Current).
3. A confirmation screen is presented. Select Continue to proceed with
the calibration or Return to Previous Screen.
Optional Equipment
Connecting External Devices
42iQTL Instruction Manual
9-7