4.10 Standby Mode

Entering Standby Mode
When the monitor is on, and Main Menu > System Setup > General Setup > Standby Mode is
set to ON, press the ON/OFF button, the dialog box displays: Sure to enter standby mode?
(Press any key to return). Select YES to enter the standby mode.
NOTE:
When the SYSTEM SETUP > STANDBY is OFF, or the monitor is in DEMO mode, or
there is any inputting signal, pressing the ON/OFF button for more than 2 seconds can
not make the monitor enter the sleep mode.
In the following two conditons, the monitor can not enter standby mode.
1. If the monitor is measuring, press the ON/OFF button for less than 2 seconds, the dialog box
displays: Cannot enter standby mode while measurement going...
2. If the battery is low, press the ON/OFF button for less than 2 seconds, the dialog box displays:
Cannot enter standby mode due to BATTERY LOW alarm!
Quitting Standby Mode
In Standby Mode, press any button on the front panel to quit standby mode.
NOTE:
1
In the following situations, the monitor will return to normal monitoring mode
automatically: The monitor receives physiological signal of SpO
the monitor is powered by battery, when the battery electric energy is low, it will enter
normal monitoring mode, and indicates low battery alarm.
2
In DEMO mode, the monitor can not enter standby mode.
- 33 -
System Menu
, and lasts for 5 s; If