Settings - Thermo Scientific RadEye GX Operating Instructions Manual

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

3.1.9

Settings

In this submenu it is possible to select between audible single
pulse indication and finder mode. A "" shows the active config-
uration. With "Autosend" the RadEye sends a data string periodi-
cally to the infrared port. "Batt.type" is used to set the correct
threshold for warning message "LOW BATT" for accumulators
(NiMH) and primary cells (Alkaline). With "Alpha Sound" and
"Alpha LED" activated each alpha pulse generates an audible
resp. optical indication. With "Set Date/Time" the user can set the
date of the real time clock. "Display αβ" shows the count rate
inclusive α rate. With "Display β" the α rate is subtracted. "Set
HV" result to another submenu where the high voltage can be set
manually or via automated plateau measurement.
Edit Tau means the you may set a fixed time constant from 1..60s
for operation mode "Ratemeter Tau".
"Language" is used to change the menu language of the RadEye
between English, German and French.
Set the display contrast with menu item "Contrast".
RadEye GX, GX-L
DB-074 E
3-11
Pi/Ff 14.09.2016