Thermo Scientific RadEye GX Operating Instructions Manual page 79

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

correction only. For individual calibration and linearity check of
dose rate detectors, the PC-software CalRadEye.EXE should be
used. Please note that the cable type and length have an influence
on the calibration parameters.
Typical parameters
Detector
HP-270
HV [V]
950
cal factor
50
[µR/h/cps]
DT [µs]
95
QC [ns]
-3.1
Over range
50,000
[cps] *)
Over range
180
power
Area [cm²]
0
Timeout [s]
100
*) Over range [cps] for RadEye GX-L is limited to 10,000 cps.
RadEye GX, GX-L
Pi/Ff 14.09.2016
HP-210
HP-
190A
900
900
20
31
50
170
-0.6
-9.5
30,000
20,000
2,000
250
20
7
60
60
DB-074 E
HP-360
EP-15
900
500
20
19
50
70
-0.6
0
30,000
30,000
2000
300
20
20
60
60
8-15