Thermo Scientific RadEye GX Operating Instructions Manual page 106

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

Working
temperature:
Storage
temperature:
Ambient air
pressure:
Relative humidity:
Operating voltage:
Degree of
protection:
EMC:
Size:
Weight:
Internal memory:
Averaging filters:
15-2
Pi/Ff 14.09.2016
-20°C ... + 50°C
-25°C ... + 50°C
700 hPa ... 1100 hPa
10 ... 90 % at 35°C not condensing
1,8 ... 4 V,
Battery low voltage starting from 2.3 V
IP 32 according to EN 60 529
Disturbance emission : EN 61000-6-3
Immunity : EN 61000-6-2
110 mm x 67 mm x 62 mm
With rubber protection, without cable
around 150g including AAA cells and pro-
tection sleeve
The last 1500 measured values are saved and
can be read out via PC program. Max- and
mean value of count rate and dose rate. The
time interval is factory preset to 120s by
default.
Logbook with 250 entries for changes of
configuration, occurring alarms and errors.
Ratemeter ADF (Advanced Digital Filter)
Digital RC-Filter with time constant
1s.....180s, depending on dose rate and dose
rate changing.
Ratemeter Tau: Linear Ratemeter with ad-
justable time constant from 1s to 60s
DB-074 E
RadEye GX, GX-L