Alarm Threshold Based On The Relative Standard Deviation (Sigma Threshold) - Thermo Scientific RadEye GX Operating Instructions Manual

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

3.6
Alarm threshold based on the relative standard
deviation (sigma threshold)
The alarm thresholds refer to the basic display set. If the sigma
value set ranges between 2 and 9, the first alarm threshold is cal-
culated from the count rate of the background ratemeter according
to the following empirical equation:
1
*
As
Cnt
n
where As1:
Cnt:
n:
T:
Av:
Count rate
Measuring
time
0...<8 cps
20 s
8...<32 cps
15s
32...
10s
<256 cps
>256 cps
5s
All count
Preset time
rates
constant
(1..60s)
3-18
Pi/Ff 14.09.2016
*
Cnt
T
Av
T
Alarm threshold 1
Count rate of the background ratemeter
sigma value set between 2 and 9
Average measuring time depending on
back ground value
Additional value (0 if linear ratemeter is enabled)
Additional
value
80
80
80
80
0
DB-074 E
ADF
Linear
ratemeter
x
x
x
x
RadEye GX, GX-L
x