Summary of Contents for Thermo Scientific Russell RL060P

  • Page 1 Russell RL060P User Guide...
  • Page 2 ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific. AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS Ultra, Sure-Flow, TEA Analyzer, Titrator PLUS, TURBO2 and Wine Master are registered trademarks of Thermo Fisher.
  • Page 3: Table Of Contents

    Table of Contents Chapter I Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Chapter II Getting .Started .
  • Page 5: Chapter I

    .card . .Some .models .also .include .a .rubber .armor . . T o .order .other .accessories . and .buffer .calibration .or .standard .solutions, .please .refer .to .please .refer .to .the . Ordering Information .section .for .more .information . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 6: Getting Started

    .3 .1/2-digit .segments .and .operation .annunciators . for .pH, .mV .or .°C .(temperature) . .Other .annunciators . include . “ HO” .(when .the .HOLD .function .is . activated) .and . “ LO” .(low .battery .condition) . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 7: Inserting New Batteries

    .the .direction .of .the .arrow .and .lift .up .the .cover . . Note .the .polarity .of .the .battery .before .inserting .it . into .position . .After .replacement, .place .the .cover . back .and .press .down .until .it .locks .tight . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 8: Battery Replacement

    .be .done .by .soaking .it .into .a .container .filled .with .pH .4 .buffer .solution .for .at .least . . 1 .hour . .Rinse .with .water .before .proceeding .to .calibrate .the .electrode .with .the .meter . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 9: Switching The Meter On

    3 . . . The .LCD .displays . “ Or” .if .the .meter .reading . exceeds .the .maximum .or . “ Ur” .if .under .minimum . measurement .range . .Refer .to .Chapter VII, Specification . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 10: Calibration

    .too .dry .or .has .not .been .used .for .a .long .period .of .time . .Always .rinse .the .probes .with . tap .water .or .rinse .solution .before .and .after .each .calibration/sample .measurement . to .avoid .cross-contamination . .For .details .refer .to .Chapter V, Electrode Care and Maintenance . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 11: Selection Of Ph Buffer Standards

    .to .abort .this .operation .if .you .do .not .wish .to .reset . . 3 . . Press . .to .confirm . . T he .meter .automatically .clears .all .stored .pH .calibration .or . mV .offset .values .and .reverts .to .measurement .mode . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 12 .mode . . 6 . . For .a .two .or .three .point .calibration, .repeat .steps .3 .through .5 .with .other .pH .buffer . values .of .4 .01 .and/or .10 .01 .for .higher .accuracy . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 13 6 . . For .a .two .or .three .point .calibration, .repeat .steps .3 .through .5 .with .other .pH .buffer . values .of .4 .01 .and/or .9 .18 .for .higher .accuracy . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 14: Ph Calibration Using Pb Standard Buffers

    .and .the .LCD .displays . “ CO” .momentarily . . T he .meter .reverts .to . measurement .mode . . 6 . . For .a .two .point .calibration, .repeat .steps .3 .through .5 .with .pH .4 .10 .buffer .for .better . accuracy . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 15 .loops .back .to .the .minimum .allowable .value . . .to .confirm .calibration . . T he .display .shows . “ CO” .momentarily .and . 6 . . Press . meter .reverts .to .measurement .mode .showing .the .current .set .value . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 16: Temperature Calibration

    .the .meter‘s .non-volatile .memory . . T o .confirm .calibration, .press . . . T he .LCD . displays . “ CO” .momentarily, .and .the .meter .reverts .to .measurement .mode . . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 17 .the .meter‘s .non-volatile .memory . . T o .confirm .calibration, .press . . . T he .LCD . displays . “ CO” .momentarily, .and .the .meter .reverts .to .measurement .mode . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 18: Measurement

    .again .to .deactivate .the .HOLD .function .or .to .release .your .frozen .reading . . The .meter .reverts .to .the .current .measurement .mode, .and .the . “ HO” .annunciator . disappears .from .the .LCD . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 19: Electrode Care And Maintenance

    . .Avoid .touching .the .glass .membrane .with .your .fingers . . W ash .thoroughly .in .tap . water .and .then .in .distilled .water . .Recalibrate .the .meter .after .cleaning .the .electrode . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 20: Troubleshooting

    Not .able .to .calibrate a . . .Display .freezes . a ..Release .reading .by . pressing .HOLD .key b . . .Faulty .electrode b . . .Replace .electrode Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 21: Specification

    Operating . T emperature . 0 .to .50 .°C Power .Requirements 4 .x . “ AAA” .alkaline .batteries Battery .Life 500 .hours Meter .Dim ./Weight 14 .x .7 .x .3 .5 .cm ./ .200 .g Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 22: Ordering Information

    .electrode .cleaning .solution .kit, .includes .one .bottle .each .of .A, . B, .C .and .D .cleaner, .15 .mL .beaker .and .pipette 900011 Ag/AgCl .electrode .filling .solution, .5 .x .60 .mL .(for .9142BN) * . .Dimensions .are .from .bottom .of .cap . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 23 .10 .01 .buffer .pouches, .10 .x .15 .mL .pouches . 911025 pH .10 .01 .buffer .pouches, .25 .x .15 .mL .pouches 911110 pH .rinse .solution .pouches, .15 .mL .x .10 .pouches 911125 pH .rinse .solution .pouches, .15 .mL .x .25 .pouches Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 24: Assistance And Warranty

    .your .local .authorized .dealer . .For .the .most .current .contact .information, . . visit .www .thermo .com/contactwater . . For .the .most .current .warranty .information, .visit .www .thermo .com/water . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 25: Declaration Of Conformity

    9001:2000 .by .QMI, .listed .as .File .#001911 . Place .and .date .of .issue: . ------------------------------------------ . Beverly, .MA, . . Patrick .Chiu . November .10, .2005 . . Quality .Assurance .Engineer . . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 26: Weee Compliance

    .recyclers .in .your .country, .and .information .on . T hermo .Scientific .products .which . may .assist .the .detection .of .substances .subject .to .the .RoHS .Directive .are .available .at . www .thermo .com/water . Thermo Scientific Russell RL060P pH Meter User Guide...
  • Page 27 Thermo Fisher Scientific Environmental Instruments Water Analysis Instruments 166 Cummings Center Beverly, MA 01915 USA Tel: 978-232-6000 Toll Free: 800-225-1480 Dom. Fax: 978-232-6015 Int’l. Fax: 978-232-6031 www.thermo.com/water 256761-001 Rev.A...