13. Retain the package in case you need to transport the EZ2 in the future. Refer to the Packing
the EZ2 section for more details. Using the original package minimizes the possibility of
damage during transportation of the EZ2.
4.3

Installing the EZ2

This section describes important actions that must be performed before operating the EZ2.
These actions include:
Removal of the EZ2 accessories and shipping material.

Installation of the AC power cord

Installation of external barcode scanner
Installation of the Wi-Fi adapter
Initial configuration
Camera exposure calibration
If an installation qualification/operational qualification (IQ/OQ) is required in your
laboratory setup, this service can be ordered together with the instrument. For details, please
contact QIAGEN Technical Service.
4.3.1

Removal of the EZ2 accessories and shipping materials

1. Remove the power cord, barcode scanner, and the quick-start guide from the foam packing
material on top of the EZ2.
2. Remove the USB flash drive, cartridge and sample/tip racks
3. Ensure all shipping foams, transport locks and other packing material have been removed as
described in the Unpacking the EZ2 section.
4.3.2
Installation of the AC power cord
1. Remove the power cord from the foam packing material on top of the EZ2.
Note: Only use the power cord that is supplied with the EZ2.
2. Ensure that the power switch is set to OFF.