After Run Maintenance - Qiagen EZ2 Connect User Manual

Hide thumbs Also See for EZ2 Connect:
Table of Contents

Advertisement

can also be used to perform decontamination by spraying worktable items, if they have been
removed from the instrument.
In case of nucleic acid contamination, DNA-ExitusPlus™ (AppliChem, cat. no. A7089,0100) can
be used for cleaning surfaces and submerging worktable items. DNA-ExitusPlus can also be used
to perform decontamination by spraying worktable items, if they have been removed from the
instrument. Cleaning with DNA-ExitusPlus can leave a residue on surfaces so for this reason, after
cleaning the items with DNA-ExitusPlus, it is required to clean the items with a wet cloth several
times, or rinse them with running water, until the DNA-ExitusPlus is completely removed.
6.2

After run maintenance

After run maintenance is required after each run on the EZ2.
The EZ2 must only be operated by qualified personnel who have been appropriately trained.
Servicing of the EZ2 must only be performed by QIAGEN Field Service specialists.
WARNING
Moving parts
To avoid contact with moving parts during the operation of the EZ2, the instrument
must be operated with the hood closed.
If the hood sensor or lock is not functioning correctly, contact QIAGEN Technical
Services
WARNING
Moving parts
Avoid contact with moving parts during operation of the EZ2. Under no
circumstances should hands be placed under the pipetting arm during movement.
Do not attempt to remove any plasticware from the workdeck whilst the instrument
is operating.
EZ2 Connect and EZ2 Connect Fx User Manual 09/2021
108

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ez2 connect fx90032109003220

Table of Contents