Note: If you want to use different disinfectants than the recommended ones, make sure that they
have the same composition.
If you are unsure about the suitability of the detergents or disinfectants to be used with the EZ2, do
not use them.
General cleaning of the EZ2, with the exception of the hood, can be done using mild detergents,
such as the Mikrozid
AF sensitive, or 70% ethanol. 70% ethanol should only be used on the
®
worktable. The hood should ONLY be cleaned with lint free tissue moistened with water.
WARNING
Toxic fumes
Do not use bleach to clean or disinfect the EZ2 instrument or used labware.
Bleach in contact with salts from the buffers can produce toxic fumes.
CAUTION
Damage to the instrument
Do not use spray bottles containing alcohol or disinfectant to clean surfaces of
the EZ2. Spray bottles should be used only to clean items that have been
removed from the worktable and if permitted by local laboratory operating
practices.
WARNING
Risk of fire
Do not allow cleaning fluid or decontamination agents to come into contact
with the electrical parts of the EZ2.
WARNING
Risk of electric shock
Do not open any panels on the EZ2.
Risk of personal injury and material damage
Only perform maintenance that is specifically described in this user manual.
EZ2 Connect and EZ2 Connect Fx User Manual 09/2021
106