Weekly Maintenance - Qiagen EZ2 Connect User Manual

Hide thumbs Also See for EZ2 Connect:
Table of Contents

Advertisement

6.4

Weekly maintenance

WARNING/
Risk of personal injury and material damage
CAUTION
Improper use of the EZ2 may cause personal injuries or damage to the instrument.
WARNING
Samples combining infectious agents
Some samples used with this instrument may contain infectious agents. Handle
such samples with the greatest of care and in accordance with the required safety
regulations.
Some chemicals used with the EZ2 instrument may be hazardous or may become
hazardous after completion of a purification.
Always wear safety glasses, gloves and a lab coat.
Venting for fumes and disposal of waste must be in accordance with all national,
state, and local health and safety regulations and laws.
Important: Before starting the weekly maintenance procedure, complete the Daily maintenance.
To maintain good contact between tip adapters and filter-tips, and to prevent liquid leaking from
the tips, a light application of grease should be applied to the O-rings of the tip adapters every
week.
1. On the touchscreen, tap Maintenance.
2. Tap Weekly maintenance. The date of the last procedure, and the date of the next due date of
the weekly maintenance are shown on the screen.
EZ2 Connect and EZ2 Connect Fx User Manual 09/2021
114