Figure 75. Protocol run completed screen.
4. To begin the procedure, tap Move down. The instrument lowers the piercing unit of the
pipettor head.
5. Open the hood.
6. Wipe the piercing unit using a soft tissue moistened with 70% ethanol. The piercing unit is
sharp. Two pairs of gloves are recommended.
Figure 76. Cleaning the EZ2 piercing unit.
EZ2 Connect and EZ2 Connect Fx User Manual 09/2021