Tube Vortexer 2 User Manual
6000-5021 (120 V)
6000-5022 (240 V)
QIAGEN
19300 Germantown Road
Germantown, MD 20874
USA
QIAGEN GmbH
QIAGEN Strasse 1
40724 Hilden
GERMANY
1087788EN Rev. 03
Sample to
Insight__
®
System Multi-Specimen
September 2015

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Hybrid Capture System Multi-Specimen Tube Vortexer 2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Qiagen Hybrid Capture System Multi-Specimen Tube Vortexer 2

  • Page 1 September 2015 ® Hybrid Capture System Multi-Specimen Tube Vortexer 2 User Manual 6000-5021 (120 V) 6000-5022 (240 V) QIAGEN 19300 Germantown Road Germantown, MD 20874 QIAGEN GmbH QIAGEN Strasse 1 40724 Hilden GERMANY 1087788EN Rev. 03 Sample to Insight__...
  • Page 2 (QIAGEN Group); DuraSeal™ (Diversified Biotech). Registered names, trademarks, etc., used in this document, even when not specifically marked as such, are not to be considered unprotected by law. © 2015 QIAGEN, all rights reserved. Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 3: Table Of Contents

    Contents Introduction ......................5 General information ..................5 1.1.1 Technical assistance ................5 1.1.2 Version management ................5 Intended use ....................5 Safety Information ....................6 Proper use ....................7 Electrical safety ...................9 Environment ....................9 Biological safety ..................9 Waste disposal ..................10 Symbols ....................11 Installation ......................14 Unpacking ....................14 Start up.....................14 Functional Description ..................15 Operational modes ...................16...
  • Page 4 Cleaning and decontamination ..............20 Regular maintenance ................21 Speed calibration ..................21 6.3.1 Prepare the materials ................21 6.3.2 Secure the specimen rack ..............22 6.3.3 Measure the RPM ................22 6.3.4 Results ....................23 Resetting the circuit breaker ...............23 Mounting feet replacement ................24 Servicing ....................25 Troubleshooting ....................26 Technical Data ....................28 Operating conditions.................28...
  • Page 5: Introduction

    1.1.2 Version management This document is Hybrid Capture System Multi-Specimen Tube Vortexer 2 User Manual; see the front cover of this user manual for document number and revision. 1.2 Intended use The HCS MST Vortexer 2 is intended for use only in conjunction with digene Hybrid Capture 2 (HC2 ®...
  • Page 6: Safety Information

    2 Safety Information This manual contains information about warnings and cautions that must be followed by the user to ensure safe operation of the MST Vortexer 2 and to maintain the instrument in a safe condition. WARNING The term WARNING is used to inform you about situations that could result in personal injury to you or other persons.
  • Page 7: Proper Use

    2.1 Proper use WARNING/ Risk of personal injury and material damage CAUTION Improper use of the MST Vortexer 2 may cause personal injuries to the user or damage to the instrument. The MST Vortexer 2 must only be operated by qualified personnel who have been appropriately trained.
  • Page 8 WARNING Risk of personal injury To avoid possible pinch hazard, load and unload the specimen rack with one hand on top of the rack handle. Use the other hand to engage or disengage the MST Vortexer 2 mechanism clamp. WARNING Risk of personal injury Do not immerse the MST Vortexer 2 in water or pour liquids over the instrument as electrical shock may occur.
  • Page 9: Electrical Safety

    (SDSs). These are available online in PDF format at www.qiagen.com/safety where you can find, view and print the SDS for each QIAGEN kit and kit component. For further information see the instructions for use that comes with the kit.
  • Page 10: Waste Disposal

    WARNING Risk of exposure to hazardous material Shake hazardous samples only in appropriate containment vessels. WARNING/ Risk of personal injury and material damage CAUTION Consider any laboratory equipment used for research or clinical analysis a potential biohazard that requires decontamination before reuse. To dispose of the MST Vortexer 2, follow all national, state and local health and safety regulation and laws for disposing of laboratory waste.
  • Page 11: Symbols

    2.6 Symbols The following symbols may be found on the instrument, in this user manual, or on labels associated with the instrument. Symbol Location Description On the instrument Warning, dangerous voltage On the instrument General warning sign CE mark for Europe Type plate on the instrument In vitro diagnostic medical device...
  • Page 12 Symbol Location Description Type plate on the Waste Electrical and Electronic Equipment instrument (WEEE) Type plate on the Serial number instrument Type plate on the Manufacturer instrument Instrument box label Fragile, handle with care Instrument box label Consult instructions for use Type plate on the Global Trade Item Number instrument...
  • Page 13 Symbol Location Description Front cover of this user Authorized representative in the European manual Community Catalog number Front cover of this user manual Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 14: Installation

    3.1 Unpacking Before using the MST Vortexer 2 for the first time, examine the exterior carton and the equipment itself for damage. In the event of shipping damage, contact your local QIAGEN representative or QIAGEN Technical Services. Carefully unpack the instrument and verify the contents of the package, which should...
  • Page 15: Functional Description

    4 Functional Description The MST Vortexer 2 is composed of a base (120V or 240V) and a vortexer platform. The MST Vortexer 2 is constructed of heavy-gauge steel that provides a stable base for steady operation, even at maximum speed. The vortexer platform can hold up to 4.5 kg (10 lbs) of weight.
  • Page 16: Operational Modes

    All of the operating controls for the MST Vortexer 2 are located on the front panel. The following figure shows the front panel. Motor speed dial Pulser toggle switch Power toggle switch Pulsation frequency dial 4.1 Operational modes The motor speed of the MST Vortexer 2 varies from 0–1600 revolutions per minute (RPM). It operates in a circular motion with an orbit of 0.51–0.71 cm (0.20–0.28 inches).
  • Page 17 The digene Specimen Rack is blue and is used with specimens collected in approved Hybrid Capture collection kits, as specified in the associated digene HC2 DNA test instructions for use. The Conversion Rack is silver and used with liquid cytology specimens collected in 15 ml conical tubes.
  • Page 18: General Operation

    5 General Operation 5.1 Continuous vortexing Note: The digene HC2 DNA tests only use the continuous vortexing operational mode. 1. Move the red lever to the horizontal position. 2. Position the specimen rack and lid on the vortexer platform until it fits snugly within the guides.
  • Page 19 5. Turn the pulser toggle switch to the ON position. 6. Adjust the pulsation frequency dial to the applicable setting. 7. To begin operation of the MST Vortexer 2, turn the power toggle switch to the ON position. The MST Vortexer 2 will pulsate until the power toggle switch is set to OFF. Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 20: Maintenance

    6 Maintenance If you have a problem with maintenance of the MST Vortexer 2, contact QIAGEN Technical Services. QIAGEN charges for repairs that are required due to incorrect maintenance. 6.1 Cleaning and decontamination WARNING/ Risk of personal injury and material damage...
  • Page 21: Regular Maintenance

    prepare a 1:20 mixture of bleach to water. When using household bleach, prepare a 1:10 mixture of bleach to water. 6.2 Regular maintenance WARNING Risk of personal injury Do not immerse the MST Vortexer 2 in water or pour liquids over the instrument as electrical shock may occur.
  • Page 22: Secure The Specimen Rack

    3. Apply a 3 x 3 cm (1 x 1 inch) strip of reflective tape to the front, top and right side of the vortexer platform. 4. Make sure that the optical tachometer is set for revolutions per minute (RPM). 6.3.2 Secure the specimen rack 1.
  • Page 23: Results

    1. Unplug the power cord from the power source. 2. Press the white button of the circuit breaker. 3. If the white button does not stay pressed, contact your local QIAGEN representative or QIAGEN Technical Services. Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 24: Mounting Feet Replacement

    6.5 Mounting feet replacement 1. Unplug the power cord from the power source. 2. Turn the MST Vortexer 2 on its side using the two handles (right and left) such that the bottom is visible and easily accessible. 3. Using a ratchet and ½ inch socket, remove the center bolt from each foot. 4.
  • Page 25: Servicing

    Do not attempt to repair the instrument. Removing the case will nullify the warranty. In the event that the product is inoperable, please contact your local QIAGEN representative and provide full failure details. When making your call, please make sure that you have the serial number of the instrument.
  • Page 26: Troubleshooting

    7 Troubleshooting Refer to this section for error handling and troubleshooting. If the recommended steps do not resolve the problem, contact QIAGEN Technical Services for assistance. Possible problem or cause Corrective action Instrument vibrates excessively Relocate the unit to a flat, even surface.
  • Page 27 Possible problem or cause Corrective action Power switch is in the ON position but no shaking occurs The motor speed dial position is at Make sure the motor speed dial position is set correctly. If problem persists, perform speed calibration. See “Speed calibration,” page 21.
  • Page 28: Technical Data

    8 Technical Data 8.1 Operating conditions Condition Parameter Dimensions (w x d x h) 242 x 280 x 369 mm (10 x 12 x 18.5 inches) Weight 21 kg (60 lbs) Power requirements 110–120 Volts AC 60Hz 220–240 Volts AC 50Hz Power consumption 50 W Air temperature...
  • Page 29: Transport Conditions

    Condition Parameter Vortexing speed 0–1600 RPM Vortexing motion Clockwise orbit Vortexing orbit diameter 0.51–0.71 cm (0.20–0.28 inches) 8.2 Transport conditions Condition Parameter Air temperature 0–60°C in manufacturer’s package 8.3 Storage conditions Condition Parameter Air temperature 0–60°C Humidity 5–80% (relative humidity); non-condensing at 30ºC Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 30: Appendix A - Waste Electrical And Electronic Equipment (Weee)

    QIAGEN provides recycling upon request at additional cost. To recycle electronic equipment, contact your local QIAGEN sales office for the required return form. After you submit the form, QIAGEN will contact you either to request follow-up information for scheduling the collection of your electronic waste or to provide you with an individual quote.
  • Page 31: Appendix B - Warranty

    Appendix B – Warranty The MST Vortexer 2 is warranted against defects in materials and workmanship for a period of one year from the date it is shipped from the manufacturer. If notified of such defects during the warranty period, the manufacturer will, at its option, either repair or replace products that prove to be defective.
  • Page 32: Ordering Information

    Ordering Information Product Contents Cat. no. 6000-5021 Hybrid Capture System 120 volt vortexer for use with digene Multi-Specimen Tube Hybrid Capture 2 DNA tests Vortexer 2 Hybrid Capture System 240 volt vortexer for use with digene 6000-5022 Multi-Specimen Tube Hybrid Capture 2 DNA tests Vortexer 2 Related products digene Specimen Rack and...
  • Page 33 This page intentionally left blank Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 34 This page intentionally left blank Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 35 This page intentionally left blank Hybrid Capture System MST Vortexer 2 User Manual 09/2015...
  • Page 36 Ordering www.qiagen.com/contact | Technical Support support.qiagen.com | Website www.qiagen.com Hybrid Capture System MST Vortexer 2 User Manual 09/2015...

This manual is also suitable for:

6000-50216000-5022

Table of Contents