Figure 77. Daily maintenance screen.
3. Clean the piercing unit (if this is not yet done during after run maintenance).
4. To prepare the piercing unit for cleaning tap Move down.
5. Open the hood.
6. Clean the piercing unit as recommended for After run maintenance.
7. Close the hood.
8. To return the piercing unit to home position, tap Move up.
9. Clean the worktable with 70% ethanol, and then with distilled water (if this is not yet done
during after run maintenance).
10. To allow tray removal, tap Move to back.
11. Check if the tray is clean. If necessary, clean it with 70% ethanol, and then with distilled
water.
12. Wipe the instrument racks with 70% ethanol, and then with distilled water.
13. Lightly wipe the O-rings of the tip adapters with a lint-free tissue to remove any residual
liquid.
EZ2 Connect and EZ2 Connect Fx User Manual 09/2021