The following tables represent the configuration switch settings for the modules without
N O T E
on-board LAN only.
Table 29
The LAN settings are done on the LAN Interface Card G1369B/C. Refer to the
N O T E
documentation provided with the card.

Communication Settings for RS-232C

The communication protocol used in the column compartment supports only
hardware handshake (CTS/RTR).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the column instrument
must be powered up again in order to store the values in the non-volatile
memory.
Table 30
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
1290 Infinity TCC User Manual
Setting the 8-bit Configuration Switch (without On-board) LAN
8-bit Configuration Switch (without on-board LAN)
Mode Select
1
RS-232C
0
Reserved
1
TEST/BOOT
1
Communication Settings for RS-232C Communication (without on-board LAN)
Mode
1
Select
RS-232C
0
2
3
4
1
Baudrate
0
1
RSVD
2
3
4
1
Baudrate
Hardware Information
5
6
Data
Bits
Reserved
SYS
RSVD
RSVD
5
6
Data Bits
11
7
8
Parity
FC
7
8
Parity
191