Eliminating Air During Operation; Down Occlusion Alarm - CME BodyGuard 575 Color Vision Operator's Manual

Pca infusion system, pain manager
Table of Contents

Advertisement

If using an ambulatory administration set without a drip chamber, priming
the line might be needed. Disconnect the set from patient before priming.

Eliminating Air During Operation

When air is detected in the line, the pump gives an audible alarm, the screen
shows AIR IN LINE, and the LED indicator turns from green to red.
To eliminate air in the line during pump operation, follow these steps: -
1.
Press
seconds again to return you to the Resume/Menu option screen.
2.
Check and resolve the cause of air detected in the line.
3.
Press
4.
Disconnect the line from the patient taking care to preserve sterility.
Select 'Prime' and press
5.
__
.
WARNING: Ensure the line is disconnected from the
patient before proceeding to the next step. Failure to do so
could result in the patient receiving a large air bolus.
6.
Press
display will back 1 step automatically. You may cease priming at any time
by pressing
7.
Repeat steps 4 & 5 until all air is eliminated from the line and fluid is seen
at the distal end of the set.
8.
Press
9.
Press
10.
Check the Infusion Summary if screen which shows: bag volume, volume
left and infused, is correct then press

Down Occlusion Alarm

After Down Occlusion Alarm press
Software ver. CVPCA30E_1
_
to cancel the alarm. Press and hold down
_
_
again and enter the relevant Access Code.
within 10 seconds to commence priming, otherwise the
.
to return to the Resume/Menu option screen.
_
to resume the current infusion.
.____
again to restart delivery.
.
45
for 3