Change the bag and press
6.
Confirm protocol data and start infusion.
NOTE: if using an infusion set with drip chamber, change to a new infusion bag,
but make sure that the drip chamber is half filled and there is no air in the
administration set.
NOTE: if using an ambulatory administration set without a drip chamber, priming
the line might be needed. Disconnect the set from patient before priming.

Eliminating Air During Operation

When air is detected in the line, the pump stops the current infusion, sounds an audible
alarm, the screen shows Air or Up Occlusion, and the LED indicator turns from green to
red.
To eliminate air in the line during pump operation, follow these steps:
Press
to silence the alarm. Press and hold down
1.
the menu screen.
2.
Enter the Access Code.
Select Prime and press
3.
4.
Enter the relevant Access Code.
5.
Disconnect the line from the patient; use aseptic technique.
Warning: ensure the line is disconnected from
the patient before proceeding to the next step.
Failure to do so could result in the patient
receiving air through into the bloodstream.
Press
within 10 seconds to start priming, otherwise the display will return to
6.
menu. You may stop priming at any time by pressing
BodyGuard 595 Color Vision™ Directions For Use
again.
.
10:00
Change Bag
Press
when done
for 3 seconds to return to
REGIONAL
10:00
Warning !
Disconnect Patient
Press
key to Prime
.
4 – Operation
-