To reinitialize the oscilloscope

When you need to restore the oscilloscope to a known configuration, use the
Default Setup key. If you press the Default Setup key and the oscilloscope does
not seem to be functioning properly, try cycling power. If the oscilloscope still
does not seem to function properly, use the following procedure.
1 Turn off the power to the oscilloscope.
2 Turn on the power to the oscilloscope.
3 When the Windows XP screen disappears press and release the Default
Setup key.
A key-down powerup completely reinitializes the oscilloscope, including the
configuration RAM. It does not affect saved waveforms or setups, which are
stored on the hard disk drive.
Figure 3-12
Key-Down Powerup
...press the power switch, then release the key
when the oscilloscope display appears.
Using the Oscilloscope
To reinitialize the oscilloscope
Press and hold
the default
setup key.
3-19