9.5
Connect SPECTANO 100 to the DTS1 dielectric test
sample box
To test the functionality of SPECTANO 100 without a sample cell, connect SPECTANO 100 to the
DTS1 dielectric test sample box by using the delivered standard triaxial cables as shown in the
following figure.
Figure 9-6
DANGER
Death or severe injury caused by high voltage or current
► Before handling SPECTANO 100 in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
and not longer than 20 meters to the ground terminal in the laboratory.
► Do not connect or operate SPECTANO 100 with a broken, dirty or
oxidized ground terminal or connection cables.
► Do not use power outlets without protective ground (PE).
► Only use the delivered AC power supply to power SPECTANO 100.
► Do not touch the output if the guard (inner shield of the triaxial
connection) is identical with the ground (outer shield of the triaxial
connection).
► Always ensure that no dangerous voltages are present at any touchable
part of the device or the DTS1 box.
1
For measuring on the Sample 2 connector, connect the blue-labeled cable to the Sample 2 connector.
Connecting SPECTANO 100 to the DTS1 dielectric test sample box for
measuring on the Sample 1 connector
1
2