SPECTANO 100 User Manual
Table 10-10
Test cell settings (continued)
Data
Outer diameter of sample
(d
)
2
Guard gap (g)
Electrode length (l)
1. Depends on the test cell type selected under Test cell configuration.
The following table describes the sweep settings.
Table 10-11
Sweep settings
Data
Test modes
Select a test mode
PDC to FDS conversion
Use PDC method for
frequencies below
Convert FDS from PDC
data until
Do not re-calculate stable
PDC data
Sampling points
Points generation
Start frequency
Stop frequency
FDS points per decade
54
Description
Outer diameter of the material sample
Gap of the guard ring
Length of the test cell's electrode
Description
Select the measurement mode for your test:
FDS (only): Frequency Domain Spectroscopy
(5 μHz...5 kHz)
PDC (only): Polarization and Depolarization
Current Method (20 μHz...99 mHz)
Combined: Combines the FDS and PDC
methods in one measurement
(20 μHz...5 kHz). The PDC method is used for
frequencies below the set limit by transforming
the PDC results to FDS results to reduce the
measurement time up to 75% compared to
frequency domain measurements (FDS).
1
Switching frequency between the FDS and PDC
methods
Defines the low-frequency limit of the
transformation from PDC to FDS.
Select the Do not re-calculate stable PDC data
check box if you do not want to recalculate the
PDC data when the data does not change with the
increasing time of the PDC measurement.
Select Auto to generate default frequency sweep
list.
Select Custom to specify your own frequency
sweep list.
Start frequency of frequency sweep
Stop frequency of frequency sweep
Sample points per decade when using the FDS
5
method