9.5
Connect SPECTANO 100 to the DTS1 dielectric test
sample box
DANGER
Death or severe injury caused by high voltage or current
Before handling SPECTANO 100 in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
the ground terminal in the laboratory.
Do not connect or operate SPECTANO 100 without proper grounding
connection.
Before handling SPECTANO 100 in any way, connect the test object to
ground properly.
Do not connect or operate SPECTANO 100 with a broken, dirty or oxidized
ground terminal or connection cables.
Do not use power outlets without protective ground (PE).
Only use the delivered AC power supply to power SPECTANO 100.
DANGER
Death or severe injury caused by high voltage or current
Do not touch the OUTPUT connector when the red LED is on or blinks.
OUTPUT onnecto
Do not touch the output if the guard (inner shield of the triaxial connection)
ouch the output if the g
is identical with the ground (outer shield of the triaxial connection).
al with the ground (
Always ensure that no dangerous voltages are present at any touchable
ensure that no da
part of the device or the DTS1 box.
the device or t
0 with a broken, di
0
rotective ground (PE
er supply to power
caused by high voltage
Installation
tial
2
cross-section to
ss-section to
grounding
nnect the test
37