SPECTANO 100 User Manual
10.4
Perform reference measurements
10.4.1
Measurement procedure
Reference
settings
42
To perform reference measurements:
1. Start SPECTANO Analyzer Suite.
2. In the start view, click the New test button
with the File tab selected on the ribbon. On the left of the test view,
the Navigation pane displays the steps of the test workflow.
Note: The Navigation pane is displayed independently of the tab selected on
the ribbon.
Figure 10-4
Test view: File tab
3. In the Navigation pane, click General test settings, and
then enter the settings (see Table 10-9 "General test
settings").4
4. In the Navigation pane, click Test cell settings.
Under Test cell configuration, select the test cell type, and then enter the
test cell settings (see Table 10-10 "Test cell
. The test view opens
settings").4