Test cell type selection in SPECTANO Analyzer Suite

Table
613-88
SPECTANO Analyzer Suite test cell selection according to different standards
Standards
AC method for solids:
IEC 60250-1969
ASTM D150-18
ASTM D150-11
AC method for liquids:
IEC 60247-2004
ASTM D924-15
DC method for solids:
62631-3-1-2017
ASTM D257-14
c
Vacuum capacitance
0
ε
Vacuum permittivity
0
8.8541878176 x 10
-12
F/m
l
Electrode length
g
Guard gap
d
Outer diameter of sample
2
d
Inner diameter of sample
1
Vacuum capacitance
2���� ∙ ( ���� + ���� )
����
= ����
c
0
����
0
0
ln �
����
2
1
%
SPECTANO Analyzer Suite selection
Cylindrical electrode with Guard ring
Cylindrical electrode with Guard ring
Note: The automated calculation of the
specific resistance and conductivity is
actually only implemented in the
SPECTANO Analyzer Suite for Disk
electrodes. All other DC parameter as
the polarization current, depolarization
current, electrical resistance or electrical
conductivity are also available for
Cylindrical electrodes. The automated
calculation of the specific resistance and
conductivity for Cylindrical electrodes
will be implemented in the future. For
the calculation of the specific resistance
and conductivity using Cylindrical
electrodes please perform a .csv export.