10.6.1

Numerical results

The measurement results are displayed numerically on the top of the result view workspace. You can
display the measurement results in frequency domain and in time domain. For information on how to
select the result display, see
10.6.3 "Result view controls".
The following table describes the numerical measurement results in the frequency domain.
Table 10-5
Numerical measurement results in the frequency domain
Data
1
Tanδ/Power factor
εr'
εr''
|Z|
Phase of Z
|Y|
Phase of Y
2
Cp/Cs
2
Rp/Rs
c'
c''
R
X
Set by the Loss angle result option in the Options dialog box (see Figure 10-10 "Options dialog box"
1.
2.
Set by the Impedance calculation control (see 10.6.3 "Result view controls"
The following table describes the numerical measurement results in the time domain (PDC).
Table 10-6
Numerical measurement results in the time domain (PDC)
Data
Polarization current
Depolarization current
R(t)
Description
Measured dissipation factor/power factor
Measured real part of the complex relative
permittivity
Measured imaginary part of the complex
relative permittivity
Absolute value of the measured impedance
Phase of the measured impedance
Absolute value of the measured admittance
Phase of the measured admittance
Measured parallel/series capacitance
Measured parallel/series resistance
Measured real part of the complex capacitance
Measured imaginary part of the complex
capacitance
Measured resistance
Measured reactance
Description
Measured polarization current
Measured depolarization current
Measured electrical resistance over time
SPECTANO Analyzer Suite
49