Cylindrical electrodes and other geometries
In this case, perform one reference measurement:
Cylindrical
electrodes
Other
geometries
1. In the Navigation pane, click Reference measurement(s), and then click
Graph and result table. On the ribbon, the Reference(s) tab is selected
automatically.
2. Click the Start reference button
to Reference 1.
3. The reference measurement is running, and the measurement results are
displayed in the graph view.
4. After the reference measurement has finished, the Start reference
button
displays the status performed
5. Now, you can start the correction procedure (see 10.4.2 "Correction
procedure" later in this chapter).
Note: You can perform the reference measurements before or after the
material measurements.
The Reference measurement(s) section and the automated correction
procedure is not available for other geometries. To determine the vacuum
capacitance c
read chapter 13.8 "Other geometries" "Vacuum capacitance c
0
and correction procedure".
SPECTANO Analyzer Suite
on the ribbon, and then click Start next
.
0
45