SPECTANO 100 Quick Start Guide
7.5
Connect SPECTANO 100 to the DTS1 dielectric test
sample box
28
DANGER
Death or severe injury caused by high voltage or current
► Before handling SPECTANO 100 in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
not longer than 20 meters to the ground terminal in the laboratory.
► Do not connect or operate SPECTANO 100 without proper grounding
connection.
► Before handling SPECTANO 100 in any way, connect the test object to
ground properly.
► Do not connect or operate SPECTANO 100 with a broken, dirty or oxidized
ground terminal or connection cables.
► Do not use power outlets without protective ground (PE).
► Only use the delivered AC power supply to power SPECTANO 100.
DANGER
Death or severe injury caused by high voltage or current
► Do not touch the OUTPUT connector when the red LED is on or blinks.
► Do not touch the output if the guard (inner shield of the triaxial connection)
is identical with the ground (outer shield of the triaxial connection).
► Always ensure that no dangerous voltages are present at any touchable
part of the device or the DTS1 box.
2
cross-section and