Check Of Factory Calibration; Test Of Output Signal Shape - Omicron Lab Bode 100 Detailed Functional & Calibration Check

Vector network analyzer
Hide thumbs Also See for Bode 100:
Table of Contents

Advertisement

1 Check of factory calibration

The following checks will help you to decide if it is required to send the Bode 100 to OMICRON Lab
for calibration.
Note:
It is assumed that the user manual of the Bode 100 was read and understood prior the
measurements described in this document are executed. If you do not have a manual at
hand you can download it at
analysis/bode-100/.

1.1 Test of output signal shape

Measurement Setup:
Please connect the Source output of Bode 100 to an oscilloscope
Figure 1: test signal shape - connection setup
Equipment Setup:
Please select the measurement type "Gain / Phase", select the source settings as follows and start a
continuous measurement:
Frequency:
10 MHz - Fixed
Output level: 0 dBm
Figure 2: test signal shape - measurement type setting
Smart Measurement Solutions
Smart Measurement Solutions
Detailed Functional & Calibration Check of Bode 100
https://www.omicron-lab.com/downloads/vector-network-
®
Bode 100 - Information
Page 2 of 21