6. We recommend correcting measurements to improve the accuracy of the measurement
results. The purpose of this correction is to eliminate the unknown capacitances like stray
capacitance or capacitance between connecting leads or between leads and the electrodes
and to correct any measurement offset. If you want to correct the measurement, perform the
reference measurement(s) as described in chapter 10.4 "Perform reference measurements".
Note: You can perform the reference measurements before or after the material
measurements. In both cases you have to enter the reference settings as described in
chapter 10.4 "Perform reference measurements".
7. In the Navigation pane, click Material measurement(s), and follow the instructions and
recommendations displayed in the test view.
8. In the Navigation pane, select the measurement you want to perform, and then enter the
measurement settings (see Table 10-13 "Measurement
Note: The reference and measurement settings depend on the test cell type selected in the test cell
settings and on the measurement mode selected in the sweep settings under Measurement
mode.
9. In the Navigation pane, click Graph and result table. On the ribbon, the Measurement(s) tab is
selected automatically.
Figure 10-
38
3
Test view: Measurement(s) tab
settings").7