SPECTANO 100 User Manual
12
Technical data
12.1
Output
Voltage range
Max. current
12.2
Input
FDS
input resistance
1
PDC
input resistance
2
FDS measurement current
PDC measurement current
1. Frequency Domain Spectroscopy
2. Polarization Depolarization Current (Time Domain Method)
12.3
Frequency range of
FDS mode
PDC mode
3
Combined
mode
3.
This mode combined the FDS and PDC method in one measurement.
12.4
Measurement range
Impedance
Capacitance
Tanδ
4.
Depends on chosen measurement mode, voltage and frequency.
70
±100 mV to ±200 V
peak
50 mA
peak
<2 Ω
2 kΩ
≤50 mA
peak
≤10 mA
5 µHz to 5 kHz
20 µHz to 99 mHz
20 µHz to 5 kHz
4
100 Ω to 20 TΩ
10 pF to 100 µF
≤ 3x10
-4