SPECTANO 100 User Manual
6.2

DTS1 dielectric test sample box

DANGER
Death or severe injury caused by high voltage or current
► Use the DTS1 box only together with the SPECTANO 100 test device and
the delivered accessories and cables.
► Read the safety instructions in 2 "Safety instructions" in this User Manual
before using the DTS1 box together with the SPECTANO 100 test device.
NOTICE
Equipment damage possible
► Do not supply a continuous input current > 50 mA
► SPECTANO 100 input can cause damages of the device.
► Always configure the max. test output voltage and the input test
frequency for the connected test object so that the maximum input current
will not be reached.
► For information on how to configure the output signals for testing with the
► DTS1 dielectric test sample box, see "Technical data" in the SPECTANO
100 User Manual.
Sample 2
RC circuit (for more information, see
the front panel of the DTS1 box) to
connect to the INPUT connector of
SPECTANO 100
Sample 1
to connect to the INPUT connector of
SPECTANO 100
Figure 6-3
16
Input connector (signal source input)
Max. continuous input current: 50 mA
to connect to the OUTPUT connector of
SPECTANO 100
DTS1 dielectric test sample box
. An overload at the
peak
peak