DANGER
Death or severe injury caused by high voltage or current
► Before handling SPECTANO 100 in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
not longer than 20 meters to the ground terminal in the laboratory.
► Do not connect or operate SPECTANO 100 without proper grounding
connection.
► Before handling SPECTANO 100 in any way, connect the test object to
ground properly.
► Do not connect or operate SPECTANO 100 with a broken, dirty or
oxidized ground terminal or connection cables.
► Do not use power outlets without protective ground (PE).
► Only use the delivered AC power supply to power SPECTANO 100.
Standard connection cable is the delivered high-temperature low-noise triaxial cable with a 1:1
connection of the inner conductor, the inner shield (guard) and the outer shield (ground). Connect the
red-labeled cable to the OUTPUT connector of SPECTANO 100 and the blue-labeled cable to the
INPUT connector of SPECTANO 100.
Figure 9-1
28
Standard high-temperature low-noise triaxial cable
2
cross-section and
Inner conductor
Inner shield (guard)
Outer shield (ground)