SPECTANO 100 User Manual
*
same potential
Figure 9-5
NOTICE
The safe area is only safe if the delivered cables and adapters are used.
Otherwise, this can also be a danger zone.
To connect SPECTANO 100 to a test cell without guard ring:
1. Ground SPECTANO 100 according to the safety rules (see chapter 2.2 "Five safety rules").
2. Connect the standard triaxial cable according to the color code to the SPECTANO 100 OUTPUT
and INPUT connectors:
Red-labeled cable to the OUTPUT connector
Blue-labeled cable to the INPUT connector
3. Depending on the test cell connection (BNC or 4 mm plug), connect one of the adapter cables
described earlier in this chapter with the corresponding color code to the standard triaxial cable:
Red-labeled adapter cable to the red marked standard cable
Blue-labeled adapter cable to the blue marked standard cable
4. Connect the adapter cables to the test cell without guard ring according to the schematic
diagram.
Note: If you are using the triaxial-to-4 mm plug adapter cable, be aware that the inner shield (guard) and
the outer shield (ground) are not connected.
For more information and example connections to different types of test cells, see chapter 13. "Test
cell configurations" in the SPECTANO 100 User Manual.
32
Inner conductor
Outer (ground) &
inner (guard) shield
*
Connecting SPECTANO 100 to a test cell without guard ring
Inner conductor
Barrier
Electrodes