Running the Example HCF Test
3. Set the Force Lower (Compression) Limit.
For specimen installation, select a low compression limit to reduce the chance of specimen
damage or injury. While -50 N is shown here as an example, you should use a limit that
adheres to your lab's specimen installation guidelines.
A. Enter -50 N in the Lower Limit text box (a negative number indicates compression).
Be sure to press Enter after typing in the limit.
B. Once again, select Interlock for the Action.

Zero the Force Signal

1. Select Exclusive Control.
In the Station Manager application main window, select the Exclusive Control check box.
This allows control to come from either the computer software or the handset, but not both.