2. Reset the Force Lower (Compression) Limit.
A. Now that your hands are no longer required to be near the specimen, increase the
force lower limit to a value that will allow clamping of the specimen in the upper grip.
B. Select Station Manager >Display > Station Setup > Channels > Landmark 1
(or other depending on your configuration) > Force, click the Channel Input
Signals button, and then click the Limits tab.
C. Enter -1000 N in the Lower Limit text box and ensure that Interlock is selected for
the Action.
Landmark™ Test System Operation
Running the Example HCF Test
75