MTS Systems Landmark 370 Series Operation page 81

Test system
Hide thumbs Also See for Landmark 370 Series:
Table of Contents

Advertisement

4. Set up a force waveform using the Function Generator.
This introduces a force waveform to the specimen.
A. In the Station Manager main window, click the Function Generator button.
B. Set up a force waveform using the following parameters:
5. Open a Station Manager Scope.
Click the Scope button in the Station Manager main window.
Landmark™ Test System Operation
Channel — Landmark 1 (or other depending on your configuration)
l
Control Mode — Force
l
Command Type — Cyclic
l
Target Setpoint — +500 N
l
Amplitude — 250 N
l
Frequency — 1.00 Hz
l
Wave Shape — Ramp
l
Compensator — None
l
Important: It is important to have Compensator set to None. If it is
left on, it can negate the function generator signal.
Running the Example HCF Test
81