Running the Example HCF Test
Note: The actuator may retract on its own when the HSM turns off due to the interlock
condition caused by exceeding the displacement limit.
A. In the Station Manager main window, click the Manual Command button to open the
Manual Command window.
B. Select the following parameters:
C. Using the slider control, slowly retract the actuator until you have enough clearance to
remove the specimen halves.
3. Ensure that the HSM is off.
The HSM should have turned off when the displacement limit was exceeded. However, make
sure that the HSM is off before nearing the crush zone.
4. Release the grip and remove the specimen half from each grip.
Using a tool that keeps your hands clear of the actuator, grasp the specimen half in the upper
grip and release the grip. Repeat the procedure for the lower grip. That completes this
Example test.

Recover from a Tripped Limit

Recovering from a tripped limit may consist of one of the following steps:
106
Channel: Landmark 1 (or other depending on your configuration)
l
Control Mode: Displacement
l
Enable Manual Command: Checked
l
Landmark™ Test System Operation