2. Adjust force limits so that they are just outside expected values for the HCF test.
A. From the Station Manager main window, select Display > Station Setup, and click
the Channel Input Signals button.
B. Select the Force channel.
C. Select the Limits tab.
D. Use the following parameters:

Show Station Manager Scope and Meters for the Test

1. It is common practice to use the Scope and Meters from the Station Manager application to
show what is happening on the physical equipment during the test. Ensure that the Scope and
Meters you set up earlier are open and running.
Note: Your meter readings will reflect your station and will be different from those
shown.
Landmark™ Test System Operation
Landmark 1 Force (Upper Limit, Upper Action) — 15000 N, Interlock
l
Landmark 1 Force (Lower Limit, Lower Action — (-1000) N, Interlock
l
Running the Example HCF Test
91