Running the Example HCF Test
6. Set up the Scope.
Set up a Scope to compare the command signal to the resulting feedback. In the Station
Manager main window, click the Signal Scope button. Enlarge the scope if necessary, and
then set up the scope as follows:
Y1 parameters:
Channel — Landmark 1 (or other depending on your configuration)
l
Signal — Command
l
Unit — N
l
Units/div — 100
l
Offset — 500
l
Y2 parameters:
Channel — Landmark 1 (or other depending on your configuration)
l
Signal — Force
l
Note: Another possible selection is Active Feedback. Active Feedback will
automatically plot your active channel (for example force or displacement).
Unit — N
l
Units/div — 100
l
Offset — 500
l
Time:
Trace Time — 5.0 s
l
82
Landmark™ Test System Operation