Running the Example HCF Test
A. In the Station Manager application main window, click the Meters button.
B. Click in the title bar of each Peak/Valley meter and change Meter Type to Running
Max/Min.
2. Switch to a Running Max/Min Force meter.
Also change the force Peak/Valley meters to Running Max/Min Meters to monitor history
during specimen removal.
A. In the Station Manager application main window, click the Meters button.
B. Click in the title bar of each Peak/Valley meter and change Meter Type to Running
Max/Min.

If Appropriate, Remove the Intact Specimen

Use this procedure only if the specimen is intact at the conclusion of the test run. If the specimen has
broken, see
"If Appropriate, Remove the Broken Specimen"
1. Select Exclusive Control.
2. Remove force from the specimen.
102
on page 105.
Landmark™ Test System Operation